In this guide, we will walk you through the process of setting up the contact closure input on a Blue-White FLEX-PRO peristaltic metering pump. This programming setup works for M3, M4, MD3, AA3, A4, and CD3 models.
Follow along with the video tutorial or use the step-by-step instructions below.
Tip: Tap the ▶ Step #
to expand or collapse more detailed instructions
Step-by-Step Guide
Step 1: Access the Input Setup menu
Step 1: Access the Input Setup menu
Press the 'Menu' button on the pump control panel. Scroll down to 'Input Setup' and press 'Enter'.
Step 3: Choose the desired contact input setting
Step 3: Choose the desired contact input setting
Select either 'Closed to Stop the Pump' or 'Open to Stop the Pump' and press 'Enter'. 'Open to Stop the Pump' is the more common option, meaning the pump will stop when the contact is open and run when the contact is closed.
Step 4: Set the delay timer
Step 4: Set the delay timer
Leave the delay timer at zero and press 'Enter' to return to the previous screen.
Step 5: Return to the main screen
Step 5: Return to the main screen
Scroll down to the bottom of the menu and select 'Escape' until you reach the main screen
Step 6: Put the pump in standby mode
Step 6: Put the pump in standby mode
Press the 'Play' or 'Start' button to put the pump into standby mode. The screen will turn blue.
Step 7: Test the contact closure input
Step 7: Test the contact closure input
Close the contact to check if the pump goes into a green running state and starts circulating.