Proper calibration of your Excel XR metering pump is essential for optimal performance and accuracy. In this guide, we will walk you through the calibration process step by step. Always read the instruction manual and follow safety procedures before starting.
Video: EXCEL XR - Calibration
Follow the video tutorial or refer to the steps below for calibrating your Excel XR metering pump.
Tip: Tap the โถ
to expand or collapse more detailed instructions
Step-by-Step Calibration Guide
Step 1: Stop the Pump and Prepare for Calibration
Step 1: Stop the Pump and Prepare for Calibration
Ensure the pump is stopped by pressing the start/stop button. Fill the calibration column and close the valve to the reservoir tank while opening the valve to the calibration column. Note the starting fluid level.
Step 2: Access the Calibration Menu
Step 2: Access the Calibration Menu
Press the Settings button on the pump display, navigate to the calibration icon, and press Enter.
Step 3: Start the Pump and Monitor Fluid Levels
Step 3: Start the Pump and Monitor Fluid Levels
Press the start/stop button to start the pump. Observe the fluid level decreasing in the calibration column.
Step 4: Stop the Pump and Review Calibration Settings
Step 4: Stop the Pump and Review Calibration Settings
Before the fluid level goes below the scale on the column, stop the pump. Note the displayed flow rate, total stroke count, and total volume.
Step 5: Adjust the Total Volume and Save Calibration
Step 5: Adjust the Total Volume and Save Calibration
Adjust the total volume to match the fluid amount drawn down from the calibration cylinder using the up/down arrow buttons. Save the calibration by pressing the Save button.
Step 6: Return the Pump to Service
Step 6: Return the Pump to Service
Close the calibration column valve and reopen the tank supply valve. The pump is now calibrated and ready for use.